Data Filtering
The averaging data filter calculates a average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 29
Filtering Levels
Setting Filtering
level
OFF No
filtering
ON (default)
Standard = short filtering (about 15 s moving average)
EXTENDED Extended
filtering
(default: 1 min average)
Use display/keypad to set the filtering level.
1. Open
the
MAIN MENU
by pressing any of the
▲▼◄►
arrow
buttons.
2. Select
Measuring
by pressing the
►
arrow button.
3. Select
Filtering
and press
CHANGE
to confirm your selection.
4. Select
Off/Standard/Extended
and press
SELECT
to confirm
your selection.
5. Press
EXIT
to return to the basic display.
FILT
Set the filtering level.
Syntax:
FILT
[
xxx
]
where
xxx
=
OFF, ON or EXT (default = ON)
Device Information
Use the display/keypad or the serial line to display the device
information.
Press the
INFO
button in the basic display to see the following
information:
- current sensor operation (for example, AutoCal or Purge) in progress
- present or past unacknowledged errors
- device
information
- adjustment information fed by the user
- measuring
settings