Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 85
List of Serial Commands
The
bold
text in the brackets indicates the default setting. Issue
commands by typing them on your computer and pressing the
Enter
key.
Table 17
Measurement Commands
Command Description
R
Start the Continuous Outputting
S
Stop the Continuous Outputting
INTV [
0
... 255
S
/MIN/H]
Set the Continuous Output Interval (for
RUN mode)
SEND [0 ... 99]
Output the Reading once
SMODE
[STOP
/RUN/POLL/SEND]
Set the Serial Interface mode
SERI [baud p d s]
User Port Settings (Default: 4800 E 7 1)
ADDR
[0
... 99]
Set the Transmitter Address (for POLL
Mode)
NET
View or set networking parameters for LAN
and WLAN interfaces
WLAN
View or set wireless network parameters
for WLAN interface
OPEN
[0
... 99]
Open Temporarily Connection to a POLL
Mode Device
CLOSE
Close the Temporary Connection (Back to
POLL Mode)
Table 18
Formatting Commands
Command Description
FORM
Set the output format of SEND and R
commands
FST
Add the state of AutoCal, purge and sensor
warming in connection with SEND and R
commands
FDATE
Add date to R and SEND outputs
FTIME
Add time to output to SEND and R outputs
Table 19
Data Recording Commands
Command Description
DIR
Lists trend log files
PLAY
[0 ... 21] [START END]
Output recorded data file. Start and end
times can only be specified if the data
logger module is installed. The times must
be given in the following format:
yyyy-mm-dd hh:mm:ss
DSEL
Select logged (and displayed) quantities
DELETE
Delete all data files, including the memory
of the optional data logger module
UNDELETE Recover
the
deleted files that have not
been overwritten