Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 83
User Port Connection
Use suitable serial cable between the
user port RxD, GND and TxD
screw terminals and the PC serial port.
Table 12
Default Serial Communication Settings for the User
Port
Parameter Value
Bauds 4800
Parity Even
Data bits
7
Stop bits
1
Flow control
None
1 2 3 4 5
6 7 8 9
+
–
RxD
GND
TxD
Ch1+
Ch1–
Ch2+
Ch2–
ANALOG OUTP
UTS
US
ER
PO
RT
POW
ER
RS
-2
32
10
…
36 V
=
24
V~
IOIOI
0506-033
Figure 57
Connection Example between PC Serial Port and User
Port
Connections to pins 4,6,7 and 8 are required only if your software
requires hardware handshaking.
User Port Operating Modes
When the transmitter is powered up, it will behave according to its
configured operating mode:
- In STOP mode, the transmitter outputs the software version and the
command prompt (if echo is on).
- In RUN mode a measurement output starts immediately.
- In POLL or MODBUS mode, the transmitter does not output anything
after power-up.
For a description of the modes, see section SMODE on page 117.
NOTE
RS-232 User Port cannot be used when a communication module