Table 26 MOL command
Syntax
Description
MOL [n]
<cr>
[n]
= Molar mass (in g/mol) of the gas in the
measurement environment.
Examples:
mol 146.06
Mol : 146.0600
mol
Mol : 146.0600 ?
You can also set the molar mass when you connect DMT152 to the MI70 indicator.
5.6 Serial Line Output Commands
5.6.1 Start Measurement Output
Use the
R
command to start the continuous outputting of measurement values as an ASCII text
string to the serial line. The format of the measurement message is set with the
FORM
command.
Table 27 R Command
Syntax
Description
r
<cr>
Starts continuous measurement output.
Example:
r
Tdf= -2.5 'C H2O= 4919 ppm
Tdf= -2.5 'C H2O= 4918 ppm
Tdf= -2.5 'C H2O= 4917 ppm
...
Outputting the results continues in intervals issued with the command
INTV
. You can stop the
output by pressing the
Esc
key, or with the command
S
. Since the interface is half-duplex, you
must enter the commands when the transmitter is not outputting.
DMT152 User Guide
M210865EN-J
46