![Vaisala Drycap DMT152 Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362040.webp)
Table 14 STATUS command
Syntax
Description
STATUS
<cr>
Select at the prompt (
?
) whether the alarm is
enabled or disabled and define the voltage/
current levels.
The following example shows the purge status alarm enabled for channel 1:
status
Channel 1:
Status : ON ?
Level Lo : 0.00 ?
Level Hi : 5.00 ?
Channel 2:
Status : OFF ?
Level Lo : 0.00 ?
Level Hi : 0.00 ?
Level Lo
and
Level Hi
are shared between the
STATUS
and
ALARM
commands. For example, if
you change the levels using the
ALARM
command, they will also change when viewing the
STATUS
command.
5.3.3 Malfunction Alarm
If the transmitter is malfunctioning, the channel output is set to a pre-defined level. This
overrides the normal measurement output of the channel. You can specify the level using the
AERR
command:
Table 15 AERR command
Syntax
Description
AERR
<cr>
Enter the error output level at the prompt (
?
)
for each channel.
Example:
aerr
Ch1 error out : 20.000 mA ?
Ch2 error out : 0.000 V ?
The current value is shown after the colon "
:
". To set a new value, enter it after the question
mark. The error output value must be within the valid range for the selected output mode.
The error output value is displayed only when there are minor electrical faults
such as a humidity sensor damage. When there is a severe device malfunction,
the error output value is not necessarily shown.
DMT152 User Guide
M210865EN-J
38
Содержание Drycap DMT152
Страница 1: ...M210865EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT152...
Страница 8: ...DMT152 User Guide M210865EN J 6...
Страница 68: ...DMT152 User Guide M210865EN J 66...
Страница 70: ...DMT152 User Guide M210865EN J 68...
Страница 71: ......
Страница 72: ...www vaisala com...