7.10.3 Configure the Alarm LED
Use the
ALARM
command to configure the alarm LED functionality.
ALARM
[LED] [Alarm] [Level]
<cr>
Syntax
Description
LED
Enable or disable the entire LED functionality.
Allowed values:
ON
- Enable the LED functionality.
OFF
- Turn off the LED completely.
Alarm
Enable or disable the dew point alarm.
Allowed values:
ON
- Enable the dew point alarm.
OFF
- Turn off the dew point alarm.
Level
Dew point temperature above which the alarm LED will turn red.
The LED will turn green when the measured dew point
temperature falls below this level.
Range -70 ... +60 °C T
df
.
Example (show current settings):
alarm
?
LED display : OFF
Alarm display : OFF
Setpoint Td(°C): -10.00
Example (change values line by line, enter to skip line):
alarm
LED display : OFF ? on
Alarm display : OFF ? on
Setpoint Td(°C): -10.00 ? -40
Example (change alarm level to -40 °C T
df
):
alarm
on on -40
LED display : ON
Alarm display : ON
Setpoint Td(°C): -40.00
Example (turn off the dew point alarm only):