4.6.4.2 DM240FA Duct Installation Flange
DMT143 models with a G1/2” ISO228/1 connection thread can be installed directly in the
process wall through a DM240FA duct installation flange. When the transmitter is installed
directly on the process wall or pipe, note that a closing valve may be needed on both sides of
the installed transmitter so that the sensor head can be removed from the process for
calibration or maintenance.
1
2
3
4
Figure 17 DM240FA with DMT143
1
Measured gas
2
Recommended additional hole (plugged) for T
d
field check reference measurement probe
(for example, Vaisala DM70)
3
DM240FA flange (thread G1/2” ISO228/1)
4
DMT143 transmitter
Chapter 4 – Installation
35