Syntax
Description
Example (show currently stored calibration date):
>
pass
1300
>
cdate
Calibration date : 2013-10-03
Example (set calibration date to 2014-03-10):
>
pass
1300
>
cdate
2014 03 10
Calibration date : 2014-03-10
Table 32 CTEXT command
Syntax
Description
CTEXT
<cr>
Show calibration information text.
CTEXT
[text]
<cr>
Calibration information text string, max 19
characters. If you want to use spaces in the
string, enclose the text in quotes.
Example (shows the text “Calibrated at” followed by the currently stored calibration text):
pass
1300
ctext
Calibrated at Vaisala
Example (set calibration text to “Lab_1”):
pass
1300
ctext
Lab_1
Calibrated at Lab_1
Table 33 CT command
Syntax
Description
CT
[temp]
<cr>
Perform an offset correction to the internal
temperature measurement.
temp
= correct temperature in degrees Celsius
(°C).
CT
[-reset]
<cr>
Clear the offset adjustment from internal
temperature measurement.
GMP231 User Guide
M211501EN-E
42