Chapter 6 __________________________________________________ Calibration and Adjustment
VAISALA ______________________________________________________________________ 113
MPCP1 [x/y] [z]
Example:
>mpcp1
1. Reading ? 981.2
1. Reference ? 980.0
2. Reading ? 1102.1
2. Reference ? 1100.0
3. Reading ? 1203.4
3. Reference ? 1200.0
.
.
.
8. Reading ? 1703.3
8. Reference ? 1700.0
>
Analog Output Adjustment (Ch1)
In the analog output calibration the analog output is forced to the
following values:
- current output: 2 mA and 18 mA
- voltage output: 10 % and 90 % of the range
Connect PTB330 to a calibrated current/voltage meter in order to
measure either current or voltage depending on the selected output type.
NOTE
The new multipoint corrections will always cancel the previous
corrections as well as the valid date of calibration of the transmitter.
where
x
=
ON
y
=
OFF
z
=
?