20901220_EN_VACUU·PURE 10_V1.6_060722
Commissioning (operation)
Maximum permissible cable length in the VACUU·BUS system:
30 m.
An interruption in communication with accessories or the removal
of accessories causes the vacuum pump to stop immediately
and an error message to be displayed (flashing rate: 6x)
see
chapter:
6.3.2 Error – Cause – Remedy on page 59
6.3.2 Error – Cause – Remedy on page 59
5.4.1
5.4.1 VACUU·BUS detection
VACUU·BUS detection
Perform bus scan (VACUU·BUS)
1
2
3 sec
1.
2.
1.
1. The vacuum pump is switched
on and stopped.
2.
2.
First hold down the stop button
and then also hold down the
start button for 3 seconds.
3
3.
3. The LEDs for the stop and
start buttons flash for 5
seconds.
The stop button lights up. The
bus scan is completed. The
connected accessories are
detected.