19
20901220_EN_VACUU·PURE 10_V1.6_060722
Safety information
2.4.5
2.4.5 Eliminate sources of danger
Eliminate sources of danger
Connect the tubing correctly
No inadmissible pressure must be created at the vacuum pump
outlet. If the back pressure at the outlet is inadmissibly high,
pumped media may escape
see chapter:
.
>
Always ensure that the outlet line is clear and has no back
pressure. The outlet must not be blocked, to ensure that
gases can exit freely.
>
Prevent uncontrolled overpressure (e.g., due to a locked or
blocked piping system, condensate, or clogged outlet line).
>
At the gas connections, the connections for the inlet and
outlet must not be mixed up. The inlet is indicated by a
directional arrow on the connecting flange.
>
Observe the maximum pressures at the inlet and outlet of the
vacuum pump, as specified in chapter
>
The system to be evacuated as well as all hose connections
must be mechanically stable.
>
Attach hoses to optional hose nozzles (e.g., adapters
at pump connections) so that they do not come loose
unintentionally.
Danger when using regeneration mode
During regeneration mode, ambient air passes through the pump
unit. Pumped media can form reactive mixtures with ambient air.
>
Make sure that the pumped media combined with air
never leads to reactive, explosive, or otherwise dangerous
mixtures.
Avoid overpressure
Regeneration mode
Содержание VACUU-PURE 10
Страница 1: ...Instructions for use OI no 20901220 Original instructions EN VACUU PURE 10 Vacuum pump...
Страница 12: ...12 20901220_EN_VACUU PURE 10_V1 6_060722 Introduction...
Страница 30: ...30 20901220_EN_VACUU PURE 10_V1 6_060722 Safety information...
Страница 83: ...83 20901220_EN_VACUU PURE 10_V1 6_060722 Appendix 8 7 8 7 CU Certificate CU Certificate CU Certificate...
Страница 84: ...84 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 85: ...85 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 86: ...86 20901220_EN_VACUU PURE 10_V1 6_060722...
Страница 87: ...87 20901220_EN_VACUU PURE 10_V1 6_060722...