7.1 Information on maintenance work
Information on maintenance work
Motor bearings have a typical durability of 40000 operating hours.
Motor capacitors have a typical durability in the range of 10000
to 40000 operating hours depending strongly on operation condi
‑
tions including ambient temperature, humidity or load.
WARNING
Overaged motor capacitors may get hot, melt or
emit a darting flame.
An overaged capacitor may get hot or even melt. Rarely
it may emit a darting flame which could be dangerous
for personnel and equipment in the vicinity.
>
Check the capacitors in the terminal box of the vac
‑
uum pump regularly.
>
Measure the capacity of the capacitors and estimate
their operating hours.
>
Replace overaged capacitors. The capacitors have to
be replaced by an qualified electrician.
Recommended maintenance activities
Maintenance intervals
Clean surfaces
if required
Clean fan grill
if required
`
Oil change
if required
`
Replace motor capacitors
after 10000 – 40000 operating hours or
in case of decreasing capacity
Check motor
capacitors
Содержание RE 2.5
Страница 1: ...Instructions for use OI no 20901559 Original instructions EN RE 2 5 RZ 2 5 RE 6 RZ 6 Rotary Vane Pump ...
Страница 24: ...24 20901559_EN_RE_RZ_V7_230523 Introduction ...
Страница 40: ...40 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Страница 48: ...48 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Страница 73: ...73 20901559_EN_RE_RZ_V7_230523 Appendix 8 8 8 8 CU Certificate CU Certificate CU Certificate ...
Страница 76: ...76 20901559_EN_RE_RZ_V7_230523 ...
Страница 77: ...77 20901559_EN_RE_RZ_V7_230523 ...
Страница 78: ...78 20901559_EN_RE_RZ_V7_230523 ...
Страница 79: ...79 20901559_EN_RE_RZ_V7_230523 ...