60
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Cleaning and maintenance
7.1
Information on service work
Recommended maintenance intervals
Maintenance intervals*
If required
15000 h
Replace the diaphragms
x
Replace the valves
x
Replace O‑rings
x
Clean or
replace molded PTFE-hose
x
Replace pressure relief valve on EK
x
Cleaning Pumping unit
x
* Recommended maintenance interval after hours of operation and under nor-
mal operating conditions; depending on the environment and area of applica-
tion, we advise performing cleaning and maintenance as needed.
Recommended aids
1
3
2
No. Auxiliary materials
1
Round bottom flask stand
2
Protective gloves
3
Chemically‑resistant funnel
>
Always wear your personal protective equipment when per-
forming activities which may bring you into contact with haz
-
ardous substances.
Example
Recommended aids
for cleaning and
maintenance
IMPORTANT!