In liquids, direct the device as perpendicular as possible to the
medium surface to achieve optimum measurement results.
Fig. 16: Alignment in liquids
If there are agitators in the vessel, a false signal suppression should
be carried out with the agitators in motion. This ensures that the
interfering reflections from the agitators are saved with the blades in
different positions.
Fig. 17: Agitators
Through the action of filling, stirring and other processes in the ves-
sel, compact foams which considerably damp the emitted signals
may form on the medium surface.
Note:
If foams lead to measurement errors, you should use the biggest
possible radar antennas or as an alternative, sensors with guided
radar.
Orientation
Agitators
Foam generation