The use of the UVIDOC HD6 involves ultraviolet (UV) illumination. Proper precautions must be
taken to avoid eye and skin exposure to the UV light. This instrument is meant for use only by
specialised personnel that know the health risks associated with UV radiation and the chemicals
that are normally used with this instrument.
Installing the hardware
Environmental conditions
Placement
Place the UVIDOC HD6 on a level-working surface.
The imager requires a stable laboratory bench or table capable of supporting 100 kg.
The UVIDOC HD6 is intended for indoor use only.
Free space
LCD panel operation free space required: 50 cm
Right, left, rear and top minimum free space required: 10 cm
Other conditions
Do not install the equipment near a window and avoid direct sunlight. We recommend attached
blinds to nearby windows
Do not install the instrument where it may get wet or flooded
Do not install the instrument in a dusty environment
Do not install the instrument in a place constantly or excessively exposed to vibration or impacts
(next to a centrifuge or to a compressor for instance)
Remove all hot or heat source near the UVIDOC HD6, specifically from the air intake.
Do not place objects near the power outlet to ensure easy access to the power cord for
disconnection in case of emergency.
Environmental
conditions
Temperature: 18°C to 25°C
Humidity: 20% to 70% RH (no dew condensation)