Chapter 4
(3/08) Configuration
-
Page 4-11
5
HVI Offsets
The
HVI
O
FFSETS
item allows you to make small adjustments to
measured values. Offsets apply to final bale data in system testing only.
These offsets do not apply to individual repetitions, calibration, module
testing, module qualification, or diagnostics.
Adjustments can be made for UH Mean Length, Uniformity Index, Short
Fiber Index (SFI), Strength, Temperature, Micronaire, Rd, +b, Trash Area
Factor and Relative Humidity.
All items, EXCEPT Trash Area are additive values. Trash Area is a
multiplier.
Note1:
the Length offset is not applied to the length value used to
calculate Uniformity.
Note2:
The Rd and +b offsets are applied BEFORE calculating color
grade.
Note3:
The Trash Area factor is applied BEFORE calculating leaf
grade.
Figure 4-8: HVI Offsets Menu
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......