USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 2
(10/07) Installation
-
Page 2-4
4
Air Requirements
1.
The system requires 100 - 150 psi (700-1034 kPa) of clean, dry,
unregulated compressed air.
a.
The air supply connected to the unit must include an overpressure
safety device to ensure that the pressure remains less than or
equal to 150 psi (1034 kPa).
2.
Use of a dryer system, air filters, and a water trap is highly
recommended.
3.
The unit is provided with an air line for connections to the facility's
air supply.
4.
Failure to supply air that is free of water, oil, or particles will
increase maintenance and
VOID THE WARRANTY!
CAUTION:
Damage caused by connecting the instrument to
improper power and air sources is not covered under the
warranty.
The following are recommended:
Air Pressure
Î
100
psi
Amount of compressed air
Î
approximately
8.5m
3
/h
Air Tubes Diameter
Î
10 mm (3/8”)
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......