To Calibrate using only 1 calibration weight (Single Point Calibration)
1. Press
PRINT to continue. The display will now show C06.
2. Press
PRINT to access the C06 channel. The display will show [C6 0].
3. The
C6 channel is used to calibrate the scale with a known weight.
4. Press
ZERO to set the value of C6 to [C6 1]. Press PRINT . The display will flash SPAN,
and then show
[######]. Enter the calibration weight value you will use (at least 10%
of max capacity you set in
C04 by using UNIT and COUNT to move the cursor left and
right, and
TARE and ZERO move the values down and up.
5. Place the calibration weight you have on the empty scale and press
6. The scale will count down from 10 to 0. Once 0 has been reached, the display will show
CALEnd.
7. Press
PRINT to continue. The display will now show C07.
8. Press
ACCUM to save and exit the setup menu
9. The scale has now been calibrated. The display will show the value of the calibration
weight on the scale,
10. If the scale does not show the value of the calibration weight, check that the feet on the
platform are not screwed in too tightly, and verify that the platform is level.
11. Unload the scale; the display should read
000000.
12. If the scale does not display
000000, check that the feet on the platform are not
screwed in too tightly, and verify that the platform is level.
To Calibrate using only multiple calibration weights (Linear Calibration)
1. Press
PRINT to continue. The display will now show C06.
2. Press
PRINT to access the C06 channel. The display will show [C6 0].
3. The
C6 channel is used to calibrate the scale with a known weight. Press ZERO to set
the value of
C6 to [C6 2]. Press PRINT. The display will flash SPAN, and then show [LnE
2].
4. Press
ZERO to enter the number of weights you want to use (2�7); the more you use,
the more accurate the calibration will be. (an example of 2 will be used)
5. Press
PRINT to set your value. The screen will flash [dbno01] then [000100]
6. Enter the lowest calibration weight value you will use (at least 10% of max capacity you
set in
C04) by using UNIT and COUNT to move the cursor left and right, and TARE and
ZERO move the values down and up.
7. Place the calibration weight you have on the empty scale and press
PRINT.
8. The scale will count down from 10 to 0. Once 0 has been reached, the display will show
he screen will flash
[dbno02] then [000200]
9. Enter the next calibration weight value you will use by using
UNIT and COUNT to move
the cursor left and right, and
TARE and ZERO move the values down and up.
10. Place the calibration weight you have on the empty scale and press
PRINT.
11. The scale will count down from 10 to 0. Once 0 has been reached, the display will show
he screen will show
CALEnd.
Then follow steps 1-12 on the single point calibration instructions