vi)
Connect Tubing Set Luer fitting to the endoscopic system.
vii)
Connect the Tubing Set fitting to CO
2
EFFICIENT ENDOSCOPIC
INSUFFLATOR.
viii)
Start the endoscopic system by turning on the necessary light and video
equipment.
ix)
Verify that the air supply from the endoscopic system is turned off.
x)
Start CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR by pressing the
FLOW STOP/RUN button.
xi)
Monitor the unit’s volume indicator during procedure
xii)
Stop the unit at conclusion of procedure.
NOTE:
If an emergency should arise whereby there is a need to terminate the
procedure, the operator should stop the CO
2
gas flow by turning off the power to
the insufflator and promptly disconnecting the Tubing Set at the CO
2
EFFICIENT
ENDOSCOPIC INSUFFLATOR
Output port.
10.8 CO
2
VOLUME DISPLAY
The Digital Volume Display reads the total volume of CO
2
passing from the
CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR to the patient. The digital display
indicates the volume of CO
2
delivered in Liters and tenths of Liters and has a range
of 0 to 999 Liters.
10.9 TERMINATING GAS FLOW & SHUT-DOWN PROCEDURES
1.
Upon completion of the diagnostic procedure, press the FLOW STOP/RUN
button if in RUN mode.
2.
Discard the entire CO
2
Endoscopic Tubing Set according to standards and
guidelines set forth by SGNA and APIC for reprocessing of water bottles used
during Gastrointestinal Endoscopy.
3.
Turn power off and disconnect the power cord from power outlet.
4.
It is recommended that the CO
2
supply tank valve be fully closed when not in
use.