EFFICIENT ENDOSCOPIC INSUFFLATOR Operator’s Manual
Page 10
Section 6.0 FRONT PANEL CONTROLS
NOTE
: The appearance of your CO
2
EFFICIENT ENDOSCOPIC INSUFFLATOR may differ slightly
from the units shown in the illustrations and photographs. (Example: buttons may be round or square).
[1] GAS SUPPLY INDICATOR
Three colors of LED’s to indicate the relative amount of gas available in a
“D” or “E”
CO
2
supply tank.
[2] VOLUME LITERS DISPLAY
Indicates the total amount of gas used since last reset.
[3] VOLUME RESET
Clears the volume display to zero.
[4] FLOW STOP/RUN
Depression starts flow and button will illuminate. CO
2
will flow in one of two
modes set by the switch. The pressure response characteristics at which CO
2
is
delivered to distend the gastrointestinal tract is a function of the clinician and
how this clinician uses CO
2
to distend the gastrointestinal tract during a
procedure. In Managed Flow Mode, the control system will automatically
transition between the high flow rate and low flow rate as required during the
procedure to compensate for external variables.
To prevent wasting gas in the Free Flow mode, when the FLOW STOP/RUN
button is pressed initially and after 150 Liters of CO
2
is continuously delivered,
the unit automatically returns to STOP mode. An audible alert, in the form of a
single chirp, will sound when the unit automatically returns to STOP mode.
Thereafter, subsequent presses of the FLOW STOP/RUN button will resume the
delivery of CO
2
for an additional 150 Liters before automatically returning to
STOP mode. To prevent wasting gas in the Managed Flow mode, when the
FLOW STOP/RUN button is pressed initially and no changes in pressure have