Universal Medical Lab Improvements Cap Track Скачать руководство пользователя страница 4

Setup Location 

1. Unpack and verify that all of the supplied equipment is present.   

2. Choose a setup location which meets the following criteria: 

a) A bench top clearance height of 34” is required in order to fully open the lid. 

b) The clearance envelope is the space around the device required for easy operation. Choose a setup 

location which will allow for a clearance envelope of at least 32” x 24”, (with the machine at the center). 

c) Proper ventilation is necessary to prevent the overheating of the machine, as well as premature 

failure. Choose an area which will allow unencumbered air 

ow, and where the temperature remains 

between 5°C (41°F) and 30°C (86°F). 

d) The Cap Track is designed to rest on the operating surface by four rubber feet. No adjustment is 

necessary for leveling the machine, however, the surface should be 

at and level. 

e) Be sure the switch at the rear of the machine is accessible, as it is the means of emergency stop.   

Initial Setup Procedure 

If any problems are found during the initial setup procedure, contact Lab Improvements at 

[email protected]   

 

1. Plug the included Line Power Cord into the back of the machine. 

2. Plug the Line Power Cord into the wall outlet. 

3. If available, plug the an ethernet cord into the back of the machine, and connect the other end to an 

internet enabled network outlet. 

4. Ensure that the switch on the back of the machine is in the ‘On’ position (indicated by the ‘I’ side being 

depressed). 

5. Open the black lid and pour in 1 bag of Lab Improvements caps (provided separately). Close the lid. 

6. Press the power button located on the front of the machine. 

7. Allow 10-15 seconds for the machine to finish booting up. 

8. Change the Manager password. See the “Manager Password” section of this manual (page 7). 

9. Enter the “SETTINGS” menu, this would be a good time to change the language (page 7) and rack 

expiration time (page 7) if needed. 

10. Complete the “Adding Racks” steps (see page 6) for as many or few (minimum 1) racks as you’ll be 

using. New racks can always be added later. 

11. Load at least 1 empty and uncapped tube into one of the nests on the circumference of the machine. 

12. Press the green “START” button on the touch screen. Noise from the hopper area may be heard, this is 

normal. The system is preparing some of the caps for use. 

13. As instructed by the system, push up one of your newly added racks to the crosshair on the right side of 

the machine. The machine should produce a beeping noise once the rack is recognized. An additional 

noise can be heard when the machine recognises a valid rack. 

14. Press “YES” when prompted to Remove Tube Data associated with the rack. 

15. Wait for the machine to complete it’s cycle. Once complete, a prompt will appear asking you to 

“Remove the Green Tubes”, press the “SKIP” button at this time. (Note: If the test tube(s) did not have 

a barcode on them, then the system will prompt claiming “Some Errors Occurred”. This is to be 

expected, press “NO” on this prompt as well.) 

16. Remove any tubes loaded in step 8.  

 

After the Cap Track has passed this procedure, it is ready for operation. 

Содержание Lab Improvements Cap Track

Страница 1: ...Cap Track CT1 User Manual 1 800 423 2767 info universalmedicalinc com...

Страница 2: ...cation 3 Initial Setup Procedure 3 Operation 4 Dealing With Errors 5 Error Types 5 Sample Recall 6 Adding Racks 6 Expired Samples 7 Manager Password 7 Rack Expiration Time 7 Change Language 7 Change T...

Страница 3: ...z Certifications CSA UL Languages English French Spanish Features Compact benchtop size and weight allows for convenient placement within the lab Simple and intuitive 7 Touchscreen interface allows fo...

Страница 4: ...being depressed 5 Open the black lid and pour in 1 bag of Lab Improvements caps provided separately Close the lid 6 Press the power button located on the front of the machine 7 Allow 10 15 seconds fo...

Страница 5: ...ed with the rack 6 Wait for the machine to complete it s cycle The machine will inform you of any system errors that may occur during operation Should a fatal error occur contact Lab Improvements at s...

Страница 6: ...ion to skip this error which will mean the tube will not be loaded into the rack and can be run through the machine again Cap Application Errors are generated when the system has failed to get a cap o...

Страница 7: ...different samples F An image of the sample s cap will appear in this box Adding Racks The Add a Rack screen can be found under DATABASE and is used to add new racks to the system A The Rack s RFID ta...

Страница 8: ...nge Password Rack Expiration Time To change the cutoff time for Expired Racks go to the SETTINGS screen and tap on the RACK CUTOFF box This will prompt a manager password Once the password has been en...

Страница 9: ...es quality motors and electrical components which are selected for high reliability and should not need servicing 5 Remove accessories before moving All cables samples and caps must be removed before...

Отзывы: