78
8
SUPPLEMENTS
78
SU
PP
LE
M
EN
TS
Chapter
8
■
Calibration setting
* For communication only
Page
No.
Item
Initial value
Setting range
Memory Protect Command
No.
1
1 Unit (torque)
2: Nm
0: mNm, 1: Ncm, 2: Nm, 3: kNm,
4: kgm, 5: kgcm, 6: gcm
N
●
61
2 (Decimal place (torque))*
2: 0.00
0: NONE, 1: 0.0, 2: 0.00, 3: 0.000
N
●
62
3 Zero calibration (torque)
0.000V
-5.000 to 5.000 [V]
N
●
63
4
5 Equivalent input
calibration (torque)
10000
-99999 to 99999
N
●
64
2
1 Increment (torque)
0: 1
0: 1, 1: 2, 2: 5, 3: 10, 4: 20, 5: 50, 6: 100
N
●
66
2 Digital zero limit (torque)
99999
0 to 99999
N
●
67
3
4
5
3
1 Unit (speed)
0: rpm
0: rpm, 1: min
-1
N
●
68
2 Increment (speed)
1
0: 1, 1: 2, 2: 5, 3: 10, 4: 20, 5: 50, 6: 100
N
●
69
3
4
5
4
1 Pulse input (speed)
0: UTM
0: UTM, 1: ENCODER
N
●
71
2 Pulse rate (speed)
4 pulse
UTM selected
1, 4 [pulse]
Encoder selected
1 to 4000 [pulse]
N
●
72
3 Minimum input
rotation speed
0: 15rpm
Pulse rate: 4 pulses
0: 15rpm, 1: 10rpm, 2: 5rpm,
3: 3rpm, 4: 2rpm
Pulse rate: 1 pulse
0: 60rpm, 1: 40rpm, 2: 20rpm,
3: 12rpm, 4: 8rpm
N
●
78
4 Rotation stop mode
0: OFF
0: OFF, 1: 2 times, 2: 4 times, 3: 8 times
N
●
79
5
5
1 Unit (power)
1: W
0: mW, 1: W, 2: kW, 3: PS, 4: HP
N
●
73
2 Decimal place (power)
0: NONE
0: NONE, 1: 0.0, 2: 0.00, 3: 0.000
N
●
74
3 Increment (power)
0: 1
0: 1, 1: 2, 2: 5, 3: 10, 4: 20, 5: 50, 6: 100
N
●
75
4
5
Содержание TM700