52
10
Appendices
52
Appendices
Chapter
10
Initialize (INITIALIZE)
This operation rewrites the internal memory contents to the factory default contents.
In this operation, zero calibration, UTM capacity, sign, minimum scale, unit, UTM type, quick stop
mode, use of encoder, direction, zero clear, ID, date and time remain unchanged; however, other
setting values can be rewritten to the factory default values.
10-3.
Password (PASSWORD)
This is a setting for maintenance and inspection.
Do not make any changes.
10-4.
Troubleshooting
Precautions
When initialize is performed, recorded data will be also deleted.
Trouble
Countermeasures
Does not indicate the expected value.
To newly connect a sensor, calibration is required.
Use the new sensor after performing calibration.
The power cannot be turned on.
The battery charge may be running out.
Connect the dedicated AC adapter included and charge the battery.
The screen fades out automatically.
The TM400 is equipped with a function to enter into the low power
consumption mode automatically if it is left without operation
(power off).
To recover from the low power consumption mode, turn on the
power again or press any key.
When not using the power off function, set the power off setting to
0 minutes.