Function Description Related to Indicated Values
37
Function Description Relate
d to Indicated Values
Chapter
5
5-1.
Indicated values
Torque
Analog signals (±5V) from UTM
Ⅱ
are displayed as torque.
Rotation speed
Pulse signals from UTM
Ⅱ
are displayed as rotation speed.
Power
Power calculated from torque and rotation speed are displayed.
Power [W] = torque [Nm] X rotation speed X
ω
(
ω
: 2
π
/60 = 0.1047198)
5-2.
Digital zero
This is a function to set the torque readings to zero.
The digital zero reset function that releases digital zero is also included.
■
Execution procedure
1.
Switch the PC application screen to the
"Waveform" tab.
2.
Click the "Digital Zero/Digital Zero Reset"
button.
This button cannot be clicked in state of standby for the trigger condition or during waveform
acquisition.
5
Function Description
Related to Indicated Values
Based on these data, torque/power attributes of the rotation speed can be
obtained.
Key points
When digital zero reset is performed, the condition before digital zero correction
is restored.