5.CALIBRATION
41
5-7-5. Test mode
This function performs error measurement using the set values from [ZERO ERROR
SET] and [SPAN ERROR SET].
From the current (old) error value and newly obtained error value, the error value is
calculated and updated using the following equation.
New error value = old error value + newly obtained error value
Operation
Press
to register the new value to your instrument.
5-7-6. Belt length
When you select BELT LENGTH option in [ERROR MEAS.SELECT], this value will
be used as a termination reference value to exit error measurement.
Pressing
toggles units between meter and millimeter.
Measurement stops when the belt travels the distance defined by this value. It is also
used for belt speed calculation when “NORMAL MODE A”, “NORMAL MODE B”,
“NORMAL A SLOW”, “PHASE MEASURE”, “ANA. NORMAL A”, or “ANA. A
SLOW” is selected for [SPEED ENTRY SELECT].
Operation
5-7-7. Cal weight density
This function specifies the weight density (kg/m) of the test chain used for error
measurements.
Operation
OK
Options: NO, ZERO MEASURE, SPAN MEASURE
MODE
FLOW CALIBRATION
→
→
TEST MODE
UNIT
Range: 0.0001 – 99999 (5 digits, arbitrary decimal point)
MODE
FLOW CALIBRATION
→
→
BELT LENGTH
Range: 0.001 – 9999 kg/m (4 digits, arbitrary decimal point)
MODE
FLOW CALIBRATION
→
→
CAL WEIGHT DENSITY