5.CALIBRATION
35
5-6-3. Equivalent calibration
Enter and store the load cell’s rated output (mV/V) and the rated capacity
value (e.g.
the value you want to be displayed).
Enter the rated capacity vale to [BALANCE WEIGHT] field.
Enter the rated output (mV/V) for the load cell.
Make sure that
indicator is off.
Operation
1
)
Press [EQUIVALENT CALIBRATION].
2
)
Enter the expected load cell output (This is the value the load cell should
output when the load you entered in BALANCE WEIGHT is place on it), and
press
.
3
)
The display changes to COMPARISON screen and
starts to flicker.
When equivalent calibration is finished, the instrument will beep and
indicator will change to
.
If an error message appears on the screen, take an appropriate measure to remove
the error and retry the calibration.
4
)
Calibration data is stored in non-volatile RAM. Wait until “NOV” indication changes
to “RUN” before you can turn off the instrument.
RUN
OK
CALS
CALS
RUN
See "TROUBLESHOOTING" , on page 157 for the details of calibration
errors.