18.Interface
149
18-4-12. Automatic Transmission Mode
1. Continuity mode
Items up to "18-4-10. Commands" P147 are specifications applicable when the
communication mode is normal (handshaking mode).
The F395 is provided with an additional continuity mode and hold point mode. Change
the mode by the key setting of the F395.
Since the setting is determined when the power is turned on, if the mode setting is
changed, turn the power off and then on again. In these two modes, no set values can be
read or written through the RS-232C.
2. Hold point mode
The following are transmitted upon completion of sampling.
・
The RHGA command is automatically output through the RS-232C.
・
When the SI/F Hold Point Transmission setting (See "18-3-6. SI/F Hold Point
Transmission" P134.) is "Transmit", the data of the hold point set under SI/F Hold
Point (See "18-3-7. SI/F Hold Point" P134.) is output in auto print format. If the hold
point set under SI/F Hold Point does not exist, automatic printing is not performed.
R A ±
Terminator
F395
0 0 0 . 0
R B ±
F395
0 0 0 . 0
R C SP
F395
0 0 0 0 0
R A ±
F395
0 0 0 . 0
R B ±
F395
0 0 0 . 0
R C SP
F395
0 0 0 0 0
Terminator
Terminator
Terminator
Terminator
Terminator
Transmission continues.