About calibration error 2 (cErr2)
How to connect a resistor to shift the zero point is as shown below.
About calibration error 3 (cErr3)
How to connect a resistor to shift the zero point is as shown below.
Connect the resistor
bEXC and -SIG.
+EXC
+SIG
-EXC
-SIG
・
This table is for 350
Ω
load cell.
・
The temperature coefficient of the connected resistor directly influences the
accuracy of the controller. Use a resistance having a temperature coefficient of
50ppm/
℃
or more (recommended value of about 5ppm/
℃
).
Resistor
Strain
Ideal
Approx.
μ
-STRAIN
mV/V
875
k
Ω
866
k
Ω
200
0.1
437
k
Ω
442
k
Ω
400
0.2
291
k
Ω
294
k
Ω
600
0.3
219
k
Ω
221
k
Ω
800
0.4
175
k
Ω
174
k
Ω
1000
0.5
146
k
Ω
147
k
Ω
1200
0.6
125
k
Ω
124
k
Ω
1400
0.7
109
k
Ω
110
k
Ω
1600
0.8
97
k
Ω
97.6
k
Ω
1800
0.9
87.3
k
Ω
86.6
k
Ω
2000
1.0
79.4
k
Ω
78.7
k
Ω
2200
1.1
72.7
k
Ω
73.2
k
Ω
2400
1.2
67.1
k
Ω
66.5
k
Ω
2600
1.3
62.3
k
Ω
61.9
k
Ω
2800
1.4
58.2
k
Ω
57.6
k
Ω
3000
1.5
54.5
k
Ω
54.9
k
Ω
3200
1.6
51.3
k
Ω
51.1
k
Ω
3400
1.7
48.4
k
Ω
48.7
k
Ω
3600
1.8
45.9
k
Ω
46.4
k
Ω
3800
1.9
43.6
k
Ω
43.2
k
Ω
4000
2.0
41.5
k
Ω
41.2
k
Ω
4200
2.1
39.6
k
Ω
39.2
k
Ω
4400
2.2
37.9
k
Ω
38.3
k
Ω
4600
2.3
36.3
k
Ω
36.5
k
Ω
4800
2.4
34.8
k
Ω
34.8
k
Ω
5000
2.5
Point
Connect the resistor
between -EXC and -SIG.
+EXC
+SIG
-EXC
-SIG
Содержание F160
Страница 1: ...01AUG2017REV 1 06 WEIGHING CONTROLLER F160 CC Link Ver 1 10 OPERATION MANUAL...
Страница 9: ...CONTENTS VIII CONTENTS VIII M E M O...
Страница 84: ...75 8 SPECIFICATIONS 75 SPECIFICATIONS Chapter 8 Equipped with BCD parallel data output interface option Unit mm...
Страница 85: ...76 8 SPECIFICATIONS 76 SPECIFICATIONS Chapter 8 Equipped with CC Link interface option Unit mm...
Страница 99: ......