4
Advanced use
71
78
:
Test Mode
This is for shipping checks.
Cautions in the maintenance mode
In the maintenance mode, normal recording is not performed.
Do not perform maintenance during recording of data.
The test mode is an item prepared for shipping checks. Do not use this mode. (It is locked by a
password.)
4-4.Over records
By this function, recorded data outside the range set by the indicator in connection or the DS252 is
identified as over status (R) and range over (U/L).
Over status (R)
If LOAD (A/D converter input over), OFL1 (net weight > 99999), or a zero alarm occurs on the
indicator side, "R" is recorded in the STATE column and the data is not added to grand totals/sub totals.
Also, it is not counted up.
Through the SI/F, the over status is automatically sent from the indicator. In case of BCD input, "over"
recording is performed when a signal is input to pin No. 46 [over input].
The over status is recorded due to an abnormality on the indicator side. The problem cannot be solved
by setting the DS252.
Setting name
78
:
Test Mode (TEST MODE)
Display
Precautions
This mode is used for factory-shipping.
It is locked by a password.
7 8 : T E S T M O D E
E X E C U T E → [ E N T ]
Note