7
Troubleshooting
103
7-2.The DS252 does not operate normally.
The DS252 is not connected to a power source.
Check that the power cords are securely fixed. The power supply voltage of the DS252
is 24V DC. Check the polarity and then connect the power supply.
Recording is not performed until the data input terminals (SI/F and/or BCD input or
RS-232C) are properly connected and recorded data is sent.
Check to see if the connectors are properly inserted and check for breaks or incorrect
wiring.
Check that the equipment sending data to the DS252 (indicator sending SI/F signals,
PC sending RS-232C signals, etc.) is in normal operation.
In case of connection by the BCD input, check that the logics of the DS252 and target
equipment are in agreement.
If the case of connection by the RS-232C, check that the communication conditions of
the DS252 and target equipment are in agreement. Settings that need to be in
agreement are the following six items (five settings): transmission speed, length of
character, stop bit, parity bit and terminator.
Recording is not performed if the setting of "Record Every ON/OFF" is OFF (invalid).
Set "Record Every ON/OFF" to ON (valid).
Normally, the DS252 records input values in the internal memory. Writing on a CF
card is performed by WRITE/GT operation.
Properly insert a CF card in the DS252 and perform WRITE/GT operation.
The DS252 does not run.
Trouble 1
Cause 1
Remedy 1
No recording is performed.
Trouble 2
Cause 1
Remedy 1
Remedy 2
Remedy 3
Remedy 4
Cause 2
Remedy 1
Cause 3
Remedy 1
Содержание DS252
Страница 1: ...PAPERLESS PRINTER DS252 OPERATION MANUAL Rev 1 03 Paperless Printer ...
Страница 85: ...4 Advanced use 74 MEMO ...
Страница 102: ...91 6 Recorded data amount 92 Data format 93 CF card 97 Recorded data examples 98 Recorded data ...
Страница 123: ...Appendix 112 External views Front view Side view Rear view Unit mm Panel cut dimensions Panel thickness 1 6 to 3 2mm ...
Страница 124: ......