Specifications
■
Sensor input section
- Sensor input for load (strain gauge input fixed)
Sensor excitation voltage
10V or 2.5V DC ±10% (factory default: 2.5V)
Output current;
within 30mA
Signal input range
-3.0mV/V to +3.0mV/V
Zero calibration range
-3.0mV/V to +3.0mV/V
Equivalent input calibration range
-3.0mV/V to -0.5mV/V
+0.5mV/V to +3.0mV/V
Equivalent input calibration error
Within 0.1% F.S.
Actual calibration range
-3.0mV/V to +3.0mV/V
(Calibration cannot be performed in the range of approx.
-0.005mV/V to +0.005mV/V from the zero calibration point.)
Accuracy
Nonlinearity;
within 0.02% F.S. ±1 digit
(at 3.0mV/V input)
Zero drift;
within 0.5
μ
V/
℃
RTI
Gain drift;
within 0.01%/
℃
Analog filter
Selectable from 10, 30, 100, and 300Hz.
Low-pass filter (-6dB/oct)
A/D converter
Rate;
4000 times/sec.
Resolution;
24 bits (binary)
Effective resolution;
approx. 1/30000 to 3.0mV/V
Analog voltage output
Output level;
approx. 2V per 1.0mV/V input
Load resistance;
2k
Ω
or more
- Sensor input for displacement (standard: pulse input (open collector))
Maximum input frequency
50kHz
Internal counting range
Approx. 1000000
Adaptable sensor
Output;
Incremental type 2-phase output
(A/B signal output)
Also capable of single-phase output
(A-phase input used. All pulses are counted as
in the plus direction.)
Output stage circuit specification;
open collector
(NPN type, Vceo=30V or more,
Ic=30mA or more)
8
SPECIFICATIONS
Содержание DeviceNet F381A
Страница 1: ...15APR2013REV 3 10 DYNAMIC FORCE PROCESSOR F381A OPERATION MANUAL ...
Страница 9: ...Contents VIII Contents VIII M E M O ...
Страница 34: ...25 2 INSTALLATION AND CONNECTION 25 INSTALLATION AND CONNECTION Chapter 2 M E M O ...
Страница 147: ...138 8 SPECIFICATIONS 138 SPECIFICATIONS Chapter 8 8 2 Outside dimensions Unit mm ...
Страница 164: ......