Service manual Phyaction V
Service manual version 0.1
Page 8
6.
Safety inspection list ( yearly check-up )
Customer identity
Unit:
Serial number:
Date:
Test 1: General inspection
Pass Fail
N.A
.
Results of previous safety inspections are available
The user manual is present
The device label and suppliers label are clearly readable
The enclosure, the intensity controls, the keys and the display are
undamaged
Mains entry and power cord are undamaged
The output connectors are undamaged
The electrode connectors and cables are undamaged.
All the operation functions work
The measured output signals are correct
All the alarm functions work.
Remarks:
Test 2: electrical safety test according to VDE 0751
Parameter Measured
Limit
Comments
Pass Fail
Protective earth resistance
<
0.2
Enclosure leakage current
µA < 1000 µA
Patient leakage current
µA < 5000 µA
The unit ……. With serial number …… has passed/failed the test
Name service engineer
Signature