Connections
18
Fig. 4.2:
Connections
1. Output process gas
7. Input process gas 3
2. Input calibration gas 1
8. Input process gas 4
3. Input process gas 1
9. Input calibration gas 1
4. Input process gas 2
10. Output condensate
5. Lead-throughs for power supply
11. Output pressure regulator leakage
6. Input ambient air
12. Input compressed air
The compressed air must be dry and free of dust; otherwise, the valves can be
destroyed.
Control pressure: 6 - 8 bar (87 - 115 psi). Every 15 minutes, 50l of compressed air
is required for 60 seconds.
4.1 Accessories
WARNING
Risk of injury/defective!
Use of non-approved accessories can cause defects and be hazardous. This will
render the warranty null and void. The operator is then liable for any damage that
may occur.
Only use original accessories or accessories that have been approved by Union
Instruments GmbH.
5
1
6
7
8
12
10
9
11
2
3
4
Содержание INCA6001
Страница 1: ...Apr 2016 V1 10 Translation of the original operating instructions Process gas analyser INCA6001 ...
Страница 3: ...3 Dimensions ...
Страница 10: ...EU Declaration of Conformity 10 ...
Страница 30: ...Transport setup and acceptance 30 ...
Страница 32: ...Startup switching on 32 ...
Страница 46: ...Operation 46 ...
Страница 48: ...Decommissioning switching off 48 ...
Страница 52: ...Maintenance 52 ...
Страница 56: ...Service 56 ...
Страница 58: ...Associated documents 58 ...
Страница 60: ...Disposal 60 ...
Страница 62: ...Spare parts 62 ...
Страница 63: ...63 16 Annex ...
Страница 64: ...Annex 64 EU Declaration of Conformity Flame arrester ...
Страница 65: ...Annex 65 ...