Troubleshooting
54
11.2 Changing/replacing fuses
Fuses may only be exchanged by an electrician or service professional. Choose
the type approved by Union.
11.3 Messages/malfunctions on the display
11.3.1 Display of messages/malfunctions
If errors occur during operation, the control system automatically switches to
overview to display priority messages.
11.3.2 Visualizing the error list
NOTE
The following list contains errors and messages that are visually displayed.
Troubleshooting measures:
Troubleshooting list!
Measures required are assigned via column [
no.].
Error text
Error message
No.
0x30D
Minimum pump pressure not reached, Sensor EC, Pressure Air
1
0x30E
Minimum pump pressure not reached, Sensor EC, Pressure Gas
2
Additional
All additional
3
11.3.3 Troubleshooting list
The following list contains causes of faults.
No.
Description
1
Inlet air filter for ambient air clogged (Fig. 4.1)
2
Process gas outlet blocked, e.g. frozen (Fig. 4.1)
Process gas inlet closed (Fig. 4.1)
- too much condensate in the line
3
Contact service
Chapter 12!
Содержание INCA5051
Страница 1: ...Apr 2016 V1 10 Translation of the original operating instructions Process gas analyser INCA5051 ...
Страница 3: ...3 Dimensions ...
Страница 10: ...EU Declaration of Conformity 10 ...
Страница 30: ...Transport setup and acceptance 30 ...
Страница 32: ...Startup switching on 32 ...
Страница 46: ...Operation 46 ...
Страница 48: ...Decommissioning switching off 48 ...
Страница 52: ...Maintenance 52 ...
Страница 56: ...Service 56 ...
Страница 58: ...Associated documents 58 ...
Страница 60: ...Disposal 60 ...
Страница 62: ...Spare parts 62 ...
Страница 63: ...63 16 Annex ...
Страница 64: ...Annex 64 EU Declaration of Conformity Flame arrester ...
Страница 65: ...Annex 65 ...