Configuration and activation
A31003-M2000-M103-2-76A9, 07/2014
3-11
Alarm and Emergency Call Configuration, Administrator Documentation
SHB_PNG-Konfig.fm
Configuration menu
8.
Confirm with
Save
.
Pull Cord Alarm
The Pull Cord Alarm can be activated optionally. It is triggered if the pull cord contact (cord) is
not detected over a defined configurable period of time, i.e. has been removed.
For more information on alarm destinations please refer to Chapter 4, "Control ports and alarm
destinations".
Procedure:
1.
Select the
Additional Features
icon in the main menu.
2.
Select
Configuration
.
3.
Enter the five-position service PIN and confirm with
OK
.
4.
Select one of the three operating modes.
5.
Select
Alarms
.
6.
Select
Pull Cord Alarm
.
7.
Change the values of the following parameters:
Table 3-8
Parameter – Pull Cord Alarm
Parameters
Operating Mode
Value Range
Standard Value
Activation
BM/AM/PNG
On/off
On
Response Time
BM/AM
1 ... 900 s
30 s
PNG
1 ... 90 s
30 s
Prealarm Time
BM/AM
0 ... 300 s
15 s
PNG
0 ... 30 s
15 s
Alarm Destination
BM
-
1
1
The alarm destination is the same in Basic Mode of the M3 Plus for all alarms and is entered under "General Settings".
112
AM/PNG
(up to 32 digits)
(blank)
8.
Confirm with
Save
.
Alarm Signal
You can define the acoustic signalling of prealarms and alarms here. The following signals start
quietly and reach their configured maximum value after a short time.
Procedure: