1.Safety Information

This operation manual includes the safety information related to the operation of the 

tester, please read them carefully before using the tester.

     Before using the Tester, please read and understand the operating manual.
     Please keep the Operating Manual properly for future use.
     Use the Tester as specified in the Operating Manual.
     Please understand in detail the Safety Information section of the manual.
     Strictly follow the Safety Information, otherwise it may cause personal injury or 

     damage to the Tester.

Danger       identifies conditions and actions that most likely pose severe or 

                   fatal hazard(s) to the users
Warning    identifies conditions and actions that may pose severe or 

                  fatal hazard(s) to the users.
Caution     identifies conditions and actions that will pose hazard(s) or

                 damage the Tester.

Safety sign "         "has 3 meanings in this manual, the user has to pay attention to 

this sign "        " during operation.

Danger

     Do not use the Tester around explosive environment, which may cause fire and

     explosion.
     Do not make any connections when the Tester is wet or the operator's hands 

     are wet.
     Do not apply any voltage beyond the rating or specified range of the Tester.
     Do not open battery compartment while testing.

Warning

     Do not use the Tester if it is damaged or metal part is exposed.
     Do not disassemble the tester randomly. If it needs repair, please contact our

     after-sales services or our agents.
     Do not change battery or open battery compartment when the Tester is wet. 

     clean it with dry soft cloth first.
     Make sure the Tester is turned off when changing battery or opening the battery

     compartment.

Caution

     Ensure test lead is inserted into the corresponding port before measurement.

     Take the battery out from the Tester if it is not used for a long time.

     Do not expose the Tester in extreme temperature and wet environment.

     Soft cloth and mild detergent should be used to clean the surface of the Tester

No abrasive and solvent should be used

     Dry the tester before storing if it is wet.

Electrical Symbols are used on the Tester or in the manual:

           danger, warning, caution                         double or reinforced insulation
          AC                                                         Grounding
          Conforms to Standards of European Union

2.Features

The Tester is designed with intelligent micro-controller chip to maintain high 

accuracy and reliability. It can measure earth resistance of earth ground devices for 

electric power installations, electrical appliances, lightening protection systems, etc, 

and also measure earth voltage(It is highly recommended not to use the Tester 

under severe outdoor environments such as raining or thunder).

  Backlight LCD;

  Battery Check; 

  Data Hold& Data Storage(up to 20 sets);

● 

 Auto Power Off;

  Offers Precision Three-Wire & Simple Two-Wire Measurements;

  “- - - - Ω” Indication for Bad Connection to C or E Terminal When Measuring 

    Earth Ground Resistance;

● 

 “OL” Overloaded Indication;

  Double-or Reinforced-Insulated Structural Design.

3.Technical Specifications Specifications

Measuring range and accuracy(under 20      5   &   75%RH):

Compliances:

      IEC  61010-1  CATIII  600V;  Pollution Degree 2

      IEC 61557-1,5

      IEC 61010-2-31

Functions

Earth Voltage

Measure Range

0V~400V(50/60Hz)

Earth

Resistance

Best Accuracy

  (1.0%+6)

   (2.0%+20) (40    range)
   (2.0%+3)(400    or 4000    range)
(Auxiliary earth resistance 500
(accuracy    5%);earth voltage   10Vac)

0.00   ~40.00

0.0    ~400.0

0    ~4000

  40

 400

4000

Measure modes:

       (1) Earth Voltage measurement: average respond
       (2) Earth Resistance  measurement: test signal frequency : around 820Hz,

current: around 3.2mA at 40Ω range.

Measuring scope (Max. deviation:±30%):

       40    Range: 5.00    ~40.00

       400    Range: 40.0    ~400.0

       4000    Range :400    ~4000

Working condition:

       Temperature:5    ~40

       Relative humidity:   80%RH(no fog)

       Altitude:   2000 m

Storage condition:

       Temperature:-20    ~60

       Relative humidity:   75%RH(no fog)

Power:

 [1.5V Alkaline Battery (AA)*6]

Overload protection:

Earth resistance ranges: 200 Vac (10 seconds)

         Earth voltage ranges: 600 Vac (30 seconds)

20MΩ between measure circuits and the casing.

Insulation impedance: 

Product size:

 160mmx70.5mmx100mm.

Product Weight: 

around 560g.

Accessories:

Green test lead (5m)

Yellow test lead (10m)

Red test lead (20m)

Auxiliary earth stakes

Simple test leads with alligator clips

1.5V Alkaline Battery (AA)

Carrying bag

English manual

1 pc

1 pc

1 pc

2 pcs

1 set

6 pcs

1 set

1 pc

4. Tester Structure&Accessories (Figure 1)

Figure 1

LCD Display

LIGHT/LOAD button

HOLD/SAVE button

TEST button

Rotary switch

Input  terminals

Standard 3-Wires test leads

Simple 2-Wires test leads

 Auxiliary earth stakes

2

3

1

4

5

6

7

8

9

Battery sign           Battery Voltage

       8.2V
7.8V~8.2V
7.4V~7.8V
7.0V~7.4V
         7V

5. Preparation before measurement

Battery Check:

 Set the rotary switch to earth voltage or earth ground 

resistance ranges, if “          ” shows, it indicates the battery is very low and 

needs to be replaced timely, otherwise the Tester may be unable to work 

normally.

5.1