background image

I. Overview

 UT501A Insulation Resistance Tester is designed with brand-new design and 
combination of massive integrated and digital circuits; it can measure insulation 
resistance, AC voltage, etc,  and enjoys high degree of accuracy, stable performance, 
easy operation and reliability. It is used for insulation resistance measurement for 
insulation materials and various kinds of electric equipments such as transformer, 
electric machines, cables, switches, electrical appliances, a ideal tool for electric 
equipment maintenance, testing and inspection.

II. Safety Information

UT501A is designed and manufactured in compliance with: IEC61010-1 CATIII 600V, 
Double Insulation and Pollution Degree 2 standards. Operate as specified in the manual, 
otherwise the protection offered by the tester would be impaired. 

 Safety Symbols 1

Danger

Identifies conditions and actions that may pose hazard to the 

users.

Warning

Alerts users to avoid electric shock.

Caution

Identifies conditions and actions that may cause damage to 

the instrument or affect accurate measurement.

      

 

 Do not measure any AC circuit with voltages above 750V

 Do not measure in flammable places. Spark may cause explosion. 
 In case that the surface of the instrument is wet or the operator's hands are wet, 

     please do not operate this instrument. 

 Do not touch the electric conduction parts of test leads when testing.
 Do not open the battery cover when testing. 
 When insulation resistance is measured, do not touch the electric wire under test. 

Danger 

Warning

 Stop using the instrument if it works abnormally. For example: the 

     equipment is damaged or metal parts are exposed.

 Operator must be careful when voltage exceeds 33Vrms, 46.7Vacrms or 70Vdc. 

     Such kinds of voltage may cause electric shock.

 Do not replace the battery in wet conditions. 
 Ensure that all test wires are connected to the test ports of tester securely. 
 When opening the battery cover, make sure that the instrument is turned off.
 Please read carefully and understand the manual before using the instrument.  
 Use the instrument as specified in the manual at any time and store the manual for 
 further reference.
 When the instrument is testing, misoperation may lead to an accident and damage 

     of equipment. 

 Caution

 Before testing insulation resistance, the circuit under test must be completely 

     discharged and completely isolated with other power circuits.

 In case the test pen is damaged and needs to be replaced, use only the test leads 
 with the same model or identical electrical specifications. 
 When the battery indicator

       

indicates that the power runs out, do not use the 

     instrument. Remove the battery and store the instrument if it is not in use for a long 
     time.

 Do not store or use this instrument in heat, humid, flammable, explosive and strong 

     electromagnetic field environments. 

Safety Symbols 2

The instrument has double insulation or reinforced insulation. 

ACV

AC Voltage

Ground

Rated Voltage

III. Technical Specification 

Accuracy

(a% r digits); calibration per year.

Ambient temperature: 

(23 5)  

Environmental humidity: 

45% to 75% 

1. Insulation Resistance Measurement

100V 

250V

500V 

1000V 

Test Range

0.00M ~5.5 G

When it is about less than 4.0M

buzzer 

outputs alarming sounds . Notes: (100V) 0.00M ~100M

Opened Voltage DC100V 10% DC250V 10% DC500V 10% DC1000V 10%

Rated Test 
Current 

When load is 
100K , the 
rated testing 
current is 0.9mA
 ~1.1mA. 

When load is 
250K , the 
rated testing 
current is 0.9mA 
~1.1mA. 

When load is 
500K , the 
rated testing 
current is 0.9mA 
~1.1mA. 

When load is 1
M , the rated 
testing current is 
0.9mA ~1.1mA. 

Shorted Current

About less than 1.8mA. 

Accuracy

0.00M ~99M

3%+5digits ; 100M ~5.5 G

5%+5digits

2. Voltage Measurement

AC Voltage 

Test Range

30V~750V(50Hz/60 Hz)

Resolution

1V

Accuracy 

2%+3digits

 Display: LCD, 1999 count.
 Low Battery Indication:

 Overload Indication: "

5.5GΩ" for insulation resistance measurement; "OL V" for 

 voltage measurement.
 Autoranging.
 Discharge voltage automatically.

 Working conditions: 0 ~35 ; relative humidity is 75% or less; altitude: 2000m.
 Storage conditions: -20 ~ 60 / relative humidity is 80% or less. 
 Dimensions: 150mm (L)  100mm (W)  71mm (D). 
 Power supply: alkaline batteries, 1.5V (5#)  6. 
 Weight: 0.5kg (with battery). 
 Accessories: test leads, instruction manual, carrying case, gallus.

 Backlight, suitable in dark site.
 Red warning light (warning light is on when high voltage outputs).

IV.  Tester's Structure (See Diagram 1)

1.  EARTH : Insulation resistance input jack.
2.  G: Negative voltage input jack.
3.  V: positive voltage input jack.
4.  LINE: Insulation resistance input jack(high voltage output)
5.  LCD screen 