OPERATING INSTRUCTIONS
USER MANUAL — 74-36105-ENG R5
31
Auxiliary power outlets and USB port (optional)
Optionally, bed can be equipped with 120V auxiliary power outlets and USB port intended to be used as convenient
power source for patient and caregiver use to energize low power devices. Power outlets are limited to 5A or
600W, with protective earth terminal. Both are located on the right side, one at the foot end of the bed, the second
under the drainage bag support. The power outlets and USB port will not be functional if bed is unplugged.
DANGER
• DO NOT use the optional auxiliary power outlets for powering a life-sustaining device.
WARNING
• The auxiliary power outlet is limited to 120V 5A supply. Ensure connected devices are within this rating.
High power device or uncertified devices may cause electrical hazards.
ATTENTION
• The organization responsable for usage and for maintenance of a medical electrical system shall ensure
that the equipments connected on the auxiliary power outlet are conforming to the electrical specifications
like, but not limited to, power drawn, leakage currents or electromagnetic compliance for electromedical
systems (IEC60601-1; 2005).
• Ensure proper power supply and proper grounding reliability with the bed power cord connected to a
hospital grade outlet. Failure to comply may cause electrical hazards, or result in loss of power to the
connected device.
• Do not connect devices in the auxiliary power outlet while their switch is turned ON.
• Do not use extension power cord, or multiple socket outlet, on the auxiliary power outlet.
• The auxiliary power outlets will not operate if bed is unplugged.
Operation:
1. Ensure that the device required to connect to the auxiliary outlet is at OFF status. Ensure that cord and
plug are in proper condition. Failure may result in inadvertent activation of the device connected and related
hazards.
2. Ensure the device has a power rating below 5A at 120VAC for a NEMA 5-15 plug, or 5V 3.1A for USB plug.
3. Insert completely the NEMA 5-15 plug or the USB plug in the auxiliary outlet. Ensure no simultaneous contacts
are made with the surroundings. Failure to comply might result in electrical shocks.
4. Ensure the bed power cord is connected to AC power through a hospital grade outlet.
5. Ensure the device to energize can be turned ON safely. Turn ON and use the device connected to the
auxiliary outlet.
!
!
!
Содержание ooksnow FL36
Страница 1: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Страница 2: ......
Страница 3: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Страница 46: ......
Страница 58: ......
Страница 74: ......
Страница 90: ......