ULVAC CRYOGENICS INCORPORATED
Disconnection of flexible hose
Disconnect flexible hoses in the opposite way of connecting. Push down the outer ring of the
UMP-40W side joint and pull out the flexible hose to disconnect.
When disconnecting flexible hoses, make sure that they return to room temperature. Note
that disconnecting hoses right after dispensing liquid nitrogen may result in damaging the
O-ring inside the coupler joint.
CAUTION