page 11 of 24

No. 47800-2-02-5

Use and operation

Installing in a vacuum system

•  Connection lines at the pump inlet have to be gas tight. Particles and dust must 

not be aspirated, the user has to provide appropriate filters if necessary. The user 

must ensure their suitability concerning gas flow, chemical resistance and safe

-

ness against clogging prior to use.

•  Connect an exhaust line gas tight at the pump outlet if necessary. Always dispose 

of exhaust gases appropriately (e.g. into a fume hood). If there is risk of release of 

dangerous or polluting fluids, install an appropriate system to catch and dispose 

of those fluids.

•  Reduce the transmission of vibration and prevent mechanical load due to rig

-

id pipelines. Insert elastic hoses or flexible elements as couplings between the 

pump and rigid pipes.

 Attention:

 Flexible elements tend to shrink when evacu-

ated.

•  The gas outlet must never be blocked. The exhaust line has always to be free 

(pressureless) to ensure an unimpeded discharge of gas.

•  Especially if the gas ballast valve is open, a power failure may cause accidental 

ventilation of the pump. In case this constitutes a potential source of danger, take 

appropriate safety measures.

• 

Pump with dual-voltage motor

: Check that the voltage changeover switch at 

the terminal box is positioned correctly.

 

Attention

: If the pump is switched on with wrong voltage selection, the motor 

may be damaged!

  Before starting the pump, check that the voltage changeover switch at the termi-

nal box is correctly positioned.

 

Change the selection at the voltage changeover switch only if the pump is 

separated from mains.

•  Make sure ventilation is adequate especially if the pump is installed in a housing 

or if the ambient temperature is elevated. Provide external ventilation if neces-

sary. Keep a distance of minimum 20 cm between fan and ambient parts.

Avoid throttling losses by using connecting pipes with large diameter and by keeping 

them as short as possible.

In case of perturbing exhaust noise connect an exhaust hose or use a silencer. Install 

outlet pipelines always falling to avoid backflow of condensate towards the pump.

Use of a suitable valve to isolate the pump from the vacuum system is recommended 

to allow the pump to warm up before pumping condensable vapours or to clean the 

pump before it is switched off.

When assembling, ensure 

vacuum-tightness

. After assembly, check the whole sys-

tem for leaks.

Secure hose connections at the pump appropriately against accidental detaching.

NOTICE

 

Voltage selection switch:

 

1. Disconnect the electrical power cord. 

 

2. Use a screw driver to adjust the

 voltage selection switch

 at the terminal  

 

box of the pump to the supply voltage: 

 

”120” corresponds to 90-126 V and ”240” corresponds to 180-253 V.

voltage selection 

switch

Содержание MD 4U NT

Страница 1: ...spect to his product page 1 of 24 No 47800 2 02 5 Diaphragm pump Instructions for use MD 4U NT The voltage changeover switch at the time of the delivery is 240 The voltage corresponds to 180 253V Appl...

Страница 2: ...age to the product Caution Hot surface Isolate equipment from mains before removing the cover NOTICE Contents Safety information 3 General information 3 Intended use 3 Setting up and installing the eq...

Страница 3: ...ring operation between 10 C and 40 C Check the maximum temperatures if installing the pump in a cabinet or a housing and make sure ventilation is adequate Install an external automatic ventilation sys...

Страница 4: ...pump without any mechanical contact except of the pump feet Comply with all applicable safety regulations Keep a distance of minimum 20 cm between fan and ambient parts e g housing walls Check fan re...

Страница 5: ...ach other and with the pump ma terials Take into consideration interactions and chemical reactions of the pumped media Ensure that the materials of the wetted parts are compatible with the pumped subs...

Страница 6: ...ntenance Due to the residual leak rate of the equipment there might be an exchange of gas albeit extremely slight between the environment and the vacuum system Adopt suitable measures to prevent conta...

Страница 7: ...ts processed by the pumping system In order to comply with law occupational health and safety regulations safety at work law and regulations for environmental protection vacuum pumps components and me...

Страница 8: ...bar 1 1 Maximum permissible pressure absolute at gas ballast valve bar 1 2 Permissible ambient temperature storage operation C 10 to 60 10 to 40 Permissible relative atmospheric moisture during operat...

Страница 9: ...ion 100 mbar low gas load 0 C to 60 C Short time 5 minutes 100 mbar low gas load 10 C to 80 C Components Wetted materials Housing cover aluminium alloy AlMgSi0 5 or AlSi12 Head cover aluminium alloy A...

Страница 10: ...be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product page 10 of 24 No 47800 2 02 5 MD 4...

Страница 11: ...ures Pump with dual voltage motor Check that the voltage changeover switch at the terminal box is positioned correctly Attention If the pump is switched on with wrong voltage selection the motor may b...

Страница 12: ...ight overheat due to insufficient cooling Do not start the pump if the pressure at outlet port exceeds maximum 1 1 bar absolute Attempts to start the pump at higher pressures may cause blockade and da...

Страница 13: ...the cap 180 gas ballast Shutdown Short term Has the pump been exposed to condensate Allow the pump to continue to run at atmospheric pressure for a few minutes Has the pump been exposed to media whic...

Страница 14: ...achieve its ultimate vacuum or usual pumping speed Centring ring at small flange connection not correctly posi tioned or leak in the pipeline or vacuum system Check pump directly con nect vacuum gauge...

Страница 15: ...ill improve the lifetime of the pump and also protect both man and environment Ensure that maintenance is done only by suitable trained and supervised technicians Ensure that the pump cannot be operat...

Страница 16: ...ed air to be blown carefully into port Remove the head cover carefully from the housing cover Note the position of the valves and remove them Replace valves if necessary Use petroleum ether or industr...

Страница 17: ...ng rod Do not mix the washers from the different pump heads Make sure that the original number is reassembled at the individual pump head Smaller number of washers The pump will not attain ulti mate v...

Страница 18: ...over Put the valves in place In case install O rings Make sure that the valves are correctly seated Valves at the outlet with round centred opening under valve valves at the inlet with kidney shaped o...

Страница 19: ...ushes 2 see figure Mount the new wire flat pin bushes and close the terminal box Fasten the cover with the four screws The replacing of the device fuse has to be carried out by an electrician Switch o...

Страница 20: ...rating conditions with every product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for...

Страница 21: ...spection for pump required No charge Immediate repairing is required Charge Periodical inspection or over haul Charge Repair order is required and estimation is required before start repairing Trouble...

Страница 22: ...um pumps Safety requirements Part2 Vacuum pumps DIN EN 61010 1 2011 IEC 61010 1 2010 Ed 3 Safety requirement for electrical equipment for measurement control and laboratory use General requirement EN...

Страница 23: ...81 45 533 0206 FAX 81 45 533 0204 Osaka Branch 3 3 31 Miyahara Yodogawa ku Osaka shi Osaka fu Japan TEL 81 6 6350 2166 FAX 81 6 6350 2169 Nagoya Branch 3 11 31 Sakae Naka ku Nagoya shi Aichi Prf Japa...

Страница 24: ...to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product page 24 of 24 No 47800 2 02 5 2...

Отзывы: