CCM Series Ceramic Capacitance Manometer
This section describes the operation of the CCMT-A series, CCMT-D series and CCMH-A
series ceramic capacitance manometers.
Before reading this section, read the standard operating procedure in the preceding sections.
17.1.
Sensor Unit Setting
Turn on the power to this unit and make sure that the following display lights for several
seconds.
For display of others, change the setting as shown below, referring to Section 6.
Display
Sensor unit
Cn3
CCMT-1000A
CCMT-1000D
CCMH-1000A
Cn2
CCMT-100A
CCMT-100D
CCMH-100A
Cn1
CCMT-10A
CCMT-10D
CCMH-10A
Cn0
-
CCMT-1D
CCMH-1A
17.2.
Front Panel Display
17.2.1.
Pressure display
State
Pressure indicator
LED
In normal measurement
0.0
×
10
+N-3
Pa to 1.3
×
10
+N
Pa
Below measurement pressure range
Display by blinking
Above measurement pressure range
FFF
When zero point adjustment is completed
0.0
×
10
+N-3
Pa
ZERO LED lights.
When zero point adjustment is reset
Pressure indication
ZERO LED lights.
When sensor unit is not connected
- - -
Err LED lights.
17.2.2.
Front Panel Operation
Inscription
Function
Remarks
PROG
Enters into the program mode
Refer to Section 5.
ZERO
Makes zero point adjustment
DEG
Not used
FIL
Not used
17.3.
Operation setting
For adjustment of the zero point, the following settings are available.
Setting
Overview
Front panel operation mode
Only operation on front panel is possible.
External I/O operation mode
Only external I/O can be operated.
RS485 communication mode
Only RS485 communication can be operated.
17.3.1.
RS425 communication operation setting “L1”, “L2”
Selects whether RS485 communication is to be used or not.
Refer to Section 22 for the detailed setting of RS485.
Display
Detail
Remarks
“L1”
Operation on the front panel or external I/O
RS485:Pressure reading only
“L2”
Only RS485 communication can be operated.
17.3.2.
Front panel and external I/O operation setting “AJ”
Selects whether execution of zero point adjustment is to be made on the front panel or by
external I/O.
Display
Detail
“F”
Operation on front panel only
“O”
Operation of I/O only