48
State
Measurement value output voltage
In normal measurement
Voltage corresponding to measured pressure
At above measurement higher limit
9.9 V or higher
At below measurement lower limit
0.25 V or less
When filament is OFF
9.9 V or higher
13.6.2.
Filament ON signal
Signal that turns on the filament. To turn on the filament, connect it to GND with the external
switch. In the case of front panel operation or RS485 communication setting, input of this signal
will be invalid.
13.6.3.
Degas ON signal
Signal that turns on degassing. To turn it on, connect it to GND with the external switch. In the
case of front panel operation or RS485 communication setting, input of this signal will be invalid.
13.6.4.
Filament 1/2 changeover signal
Filament 1/2 materials differ
Filament 1 is a yttria-coated iridium wire. Filament 2 is a tungsten wire.
For details, refer to the SH2 instruction manual.
Setting only filament1, if you used ST2
ST2 has only one filament. Setting only filament1.
Signal that changes over filament 1/2. To change it over to filament 2, connect it to GND. In
the case of front panel operation or RS485 communication setting, input of this signal will be
invalid.
1.0E-08
1.0E-07
1.0E-06
1.0E-05
1.0E-04
1.0E-03
1.0E-02
1.0E-01
1.0E+00
1.0E+01
0
1
2
3
4
5
6
7
8
9
10
PRES
SUR
E (
Pa)
OUTPUT VOLTAGE (V)
Sens
or
Err
or
Over
Ra
nge