38
12.
Hot Cathode Gauge BMR2
Cable connecting the basic unit and BMR2
The cable (24AWG) connecting the basic unit and BMR2 is 10 m long
maximum. If a longer cable is necessary, connect the power source directly
to the BMR2 or use a cable with a larger diameter.
This section describes the operation of the BMR2 ionization vacuum gauge.
Before reading this section, read the standard operating procedure in the foregoing sections.
12.1.
Sensor Unit Setting: “Sn1”
Turn on the power to this unit and make sure that “Sn1” lights for several seconds.
To display others, press the
key several times within 3 seconds after turning on power,
change the display to “Sn1” and make it definite with the
key.
For more information about the sensor unit setting, refer to Section 6.
12.2.
Front Panel
12.2.1.
Pressure display
Status
Pressure indicator
LED
In normal measurement
5.0
×
10
-8
Pa to 9.9
×
10
+0
Pa
Below measurement pressure range
0.0
×
10
-8
Pa
Above measurement pressure range and
when pressure protection signal is input
Prt
Above measurement pressure range and
when pressure protection signal is not input
FiO
When power is turned on
FiF
Immediately after filament is turned on
FiO
When degassing is ON
Pressure indication
DEG is lit
When emission valid is OFF
Pressure indication
Err blinks
When sensor unit is not connected
- - -
Err lights
*1: When a CALCULATION function is used, it displays to 10
-11
at the minimum.
10
-10
is “A (A of capital letter)”, 10
-11
is “b (B of small letter)”.Please note “b (B of small
letter)” to make a mistake as '6' of the figure.
12.2.2.
Description of front panel keys
Inscription
Function
Remarks
PROG Enters into program mode
Refer to Section 5.
ZERO Not used
DEG
Degassing ON/OFF operation
Cannot be operated when RS485 and external I/O
are set
FIL
Filament ON/OFF operation
Cannot be operated when RS485 and external I/O
are set
12.3.
Setting Operation
12.3.1.
Setting Overview
The following settings are available for turning on the filament or degassing and for changing
over the filament 1/2.
Setting
Overview
RS485 communication mode
Only RS485 communication is available.
Front panel operation mode
Only front panel operation is available.
External I/O operation mode
Only external I/O operation is available