36
11.4.
Connecting to the Sensor Unit
The table below gives the pin assignment in connecting to the sensor unit. In connecting to the
sensor unit, it is recommended to connect all the following pins.
* Connector for connection “SENSOR” : D-sub 15 pin connector male, M2.6mm screw
* Connector for connection “BPR2, SP1” : D-sub15 pin connector female, M2.6mm screw
“SENSOR”
Description
Remarks
Sensor unit
1
24 VDC power supply
24VDC power supply to the sensor unit
1
2
Burnout signal
Error signal of filament burnout is input.
2
4
Measuring unit
connection check signal
Measuring unit connection check signal
is input
4
8
Pressure signal input +
Pressure signal is input.
8
9
Power GND
Power GND to the sensor unit
9
15
Pressure signal input -
Pressure signal is input.
15
Case
FG
Frame ground
Case
11.5.
Connection with External Devices
The table below gives the pin assignment in connecting to an external device.
* Connector for connection “I/O”: D-sub 15 pin connector female, M2.6mm screw
“I/O” pin
Description
Remarks
2
Burnout signal output
Lo when error, 30VDC
MAX
, 50mA
MAX
, 70mW
3
Setpoint 1 actuating signal
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
7
Setpoint 3 actuating signal
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
8
Pressure signal
0 to 10V DC
9
Signal GND
GND of pressure signal, burnout signal, setpoint, etc.
10
RS485 -
Serial communication RS485 - output
11
Setpoint 2 actuating signal
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
12
RS485 +
Serial communication RS485 + output
15
Signal GND
GND of pressure signal, burnout signal, setpoint, etc.
Case
FG
Frame ground
11.6.
External Input/Output Signal
11.6.1.
Output voltage
Output voltage output : I/O connector 4 [8pin]
→
GND [15pin]
P=10
×
(V-E)
×
10 ^ ( E – 1 + C )
P:
pressure value (Pa)
V:
Measured value output voltage (V)
E:
Measured value output voltage from which decimal point is omitted
C:
Exponent value of Calculation value
(The value of C in A.B
×
10
C
, refer to section18)
The (V-E) calculated value may become 0.1 or lower due to errors in the sensor's output voltage
and errors with the measuring instrument. If it falls below 0.1 in this manner, we recommend
rounding up to 0.1 and calculating.
Condition
Measured output voltage
In normal measurement
Voltage corresponding to the measured pressure
When measurable higher limit is exceeded
5.1 V
When lower than measurable lower limit
0 V
When filament has burnt out
9 V or higher