![Ulvac G-Tran ISG1 Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/ulvac/g-tran-isg1/g-tran-isg1_instruction-manual_837695044.webp)
32
10.4.
Connection to the Sensor Unit: “SENSOR” Connector
The table below gives the pin assignment in the connection with the sensor unit. In the
connection with the sensor unit, it is recommended to connect all of the following pins.
* Connector for connection “SENSOR”: D-sub 15 pin connector male, M2.6mm screw
* Connector for connection “SC1” : D-sub15 pin connector female, M2.6mm screw
“SENSOR”
Description
Details
Sensor unit
1
24 VDC power
24VDC power supply to the sensor unit
1
4
Discharge check signal
*1
Signal showing that discharge : input
4
5
HV ON/OFF
HV ON/OFF signal: output
5
8
Pressure signal input +
Pressure signal: input.
8
9
Power GND
Power GND to the sensor unit
9
15
Pressure signal input -
Pressure signal: input.
15
Case
FG
Frame ground
Case
* The discharge check signal indicates that discharge current is flowing normally. The setpoint
of this instrument is not actuated unless this signal is input.
10.5.
Connection with External Devices: “I/O” connector
The table below gives the pin assignment in the connection with external devices.
* Connector for connection “I/O”: D-sub 15 pin connector female, M2.6mm screw
“I/O”
Description
Remarks
3
Setpoint 1 actuating signal output
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
4
Discharge check signal output
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
5
HV ON/OFF signal input
Actuated when shorted to GND
7
Setpoint 3 actuating signal output
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
8
Pressure signal
0 to 10 VDC
Signal GND
Signal GND
10
RS485 -
Serial communication RS485 - output
11
Setpoint 2 actuating signal output
Lo when actuated, 30VDC
MAX
, 50mA
MAX
, 70mW
12
RS485 +
Serial communication RS485 + output
15
Signal GND
Signal GND
Case
FG
Frame ground
10.6.
Signals
10.6.1.
Output voltage
Output voltage is outputted with the I/O con[8pin]
→
GND [15 pin].
P = 10
×
(V-E)
×
10 ^ ( E – 8 + C )
P : Pressure value (Pressure)
V : Measurement value output voltage (V)
E : Measurement value output voltage V from which fractions are rounded off
C : Exponent value of Calculation value
(The value of C in A.B
×
10
C
, refer to section18)
The (V-E) calculated value may become 0.1 or lower due to errors in the sensor's output voltage
and errors with the measuring instrument. If it falls below 0.1 in this manner, we recommend
rounding up to 0.1 and calculating.
Status
Measurement value output voltage
In normal measurement
Voltage corresponding to the measured pressure
Above the measurable higher limit
8.1 V or higher
Below the measurable lower limit
※
1
3 V
When HV is OFF
10 V
No discharge
10 V
※
1:Cf. instruction manual of SC1.
Содержание G-Tran ISG1
Страница 2: ......
Страница 19: ...7 2 5 Installation DIN panel Panel thickness 4mm ...
Страница 106: ...94 26 EC DECLARATION OF CONFORMITY ...
Страница 107: ...95 27 Certificate of Decontamination ...