Section 4 Equipment Installation, Piping, and Wiring
ULVAC CRYOGENICS INCORPORATED
4-7
4.6 Electrical connections
Refer to “Single cryopump installation” in previous page to connect the refrigerator cable.
Below are the connecting procedures of input power cable and remote input/output cable.
Refer to “Section 2.3 Wiring requirements” and confirm the specifications of the input
power cable and the customer-supplied power circuit at the equipment side.
Ensure that the breaker which supplies power to the compressor has been turned off on
the distribution board at the equipment side before connecting the input power cable.
Ensure that the breaker which supplies power to the compressor has been turned off on
the distribution board at the equipment side before connecting the input power cable.
To use a lock key to keep the power off status is effective as a fail safe.
To use a lock key to keep the power off status is effective as a fail safe.
4.6.1 Connecting the input power cable
4.6.1 Connecting the input power cable
Be sure to connect the input power cable first to the compressor and then to the circuit
breaker of input power supply source. Wiring connection at power supply side must be
done properly in accordance with the specifications of customer-supplied equipments.
Connect ground wire as well (D class grounding).
Be sure to connect the input power cable first to the compressor and then to the circuit
breaker of input power supply source. Wiring connection at power supply side must be
done properly in accordance with the specifications of customer-supplied equipments.
Connect ground wire as well (D class grounding).
Connect the plug of the input power cable to the compressor receptacle. After
connecting the plug to the “INPUT POWER” receptacle on the rear panel of the
compressor, turn the rotating part of the plug to the right. Turn it until it won’t turn any
more to connect the plug properly.
Connect the plug of the input power cable to the compressor receptacle. After
connecting the plug to the “INPUT POWER” receptacle on the rear panel of the
compressor, turn the rotating part of the plug to the right. Turn it until it won’t turn any
more to connect the plug properly.
Refer to “Section 2.3 Wiring requirements” and confirm the specifications of the input
power cable and the customer-supplied power circuit at the equipment side.
The electrical wiring must be done in accordance with all applicable local, national, and
international standards. Failure to observe this warning may result in damage to and/or
break down of the equipments.
CAUTION
WARNING
There is a danger of high voltage. Ensure that the input power source is OFF and locked
before connecting or disconnecting input power cable. The wiring work must be undertaken
only by an authorized electrician familiar with the structure and the risk of this equipment.
Fail re to observe this warning may result in severe bodily injury or fatal accident.
u
Содержание C30MVER
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 32: ...C30VR Compressor Instruction Manual 1 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...
Страница 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank ...