hygea

®

 6427 Daily Operation 

 
 

 

 

 

Reg. 2558071 

GB 827 1111 60 

UW-DC-112-44 

 

 

1.

 

Filling the Ultrasonic Cleaner 

Fill the 

hygea 6427

 with cold water from the tap. Ensure the drain valve is closed 

before filling. The water needs to reach the level sensor (a small screw attached 
to the inside of the tank) located in the front right hand corner. 
 
Ensure the basket and nothing else is in direct contact with the level sensor. 
 

2.

 

Adding detergent 

Add 

the correct amount of detergent (refer to manufacturer’s technical data 

sheet), if not enough detergent is added, the level sensor may not trigger if there 
is not enough conductivity in the water. 

This will display a “Fluid Level Low”.

 

 

3.

 

Powering up the Ultrasonic Cleaner 

Ensure the mains plug is in the power socket and switched on. Turn the machine 
on via the switch located on the rear of the machine, left hand side. 
 
Once the machine is turned on, it will prompt you on the LCD display to close the 
lid. Close the lid as prompted, the ultrasonics and heat will operate in order to 

heat the water to the minimum set temperature (20°C). 
 
Once the water has reached this temperature, the lid can now be opened, the 
ultrasonics will stop. The heaters will remain switched on in order to get the 
water to the set temperature (25°C). 
 

4.

 

Degassing the water 

It is recommended that one cycle is run before processing instruments, as this will 
ensure the water is degassed, and the ultrasonics perform more effectively. 
 

5.

 

Connecting instruments 

The machine is now ready for use. Connect the instruments to the manifold using 
the  accessory  kit  supplied. 

Important  note:

  when  not  processing  hollow 

instruments, ensure that at least one luerlok connector is fitted to each manifold. 
This will protect the pump from damage. 

 

6.

 

Running a cleaning cycle 

Once the instruments are in place, close the lid. To run a cycle: 

 

Press the Enter/Accept button 

 

Select  the  operator  (if  no  operators  programmed  in,  simply  choose 
OP01), by pressing the Enter/Accept button 

 

Press the Enter/Start button to begin the cycle 

 

 