SonixTouch Q+ Service Manual
00.053.204, Revision A
Appendix A: Electromagnetic Immunity Tables
A-1
APPENDIX A: ELECTROMAGNETIC IMMUNITY TABLES
Table 1-1: EN 60601-1-2:2007 (Table 1)
Table 1-2: EN 60601-1-2:2007 (Table 2)
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
The SonixTouch Q+ Ultrasound System is intended for use in the electromagnetic environment specified below. The customer or the user of the SonixTouch Q+
Ultrasound System should ensure that it is used in such an environment.
Emissions Test
Compliance Level
Electromagnetic Environment-Guidance
RF emissions CISPR 11
Group 1
The SonixTouch Q+ Ultrasound System uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class A
The SonixTouch Q+ Ultrasound System is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The SonixTouch Q+ Ultrasound System is intended for use in the electromagnetic environment specified below. The customer or the user of the SonixTouch Q+
Ultrasound System should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment-Guidance
Electrostatic
discharge (ESD)
IEC 601000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1kV for input/output lines
±2 kV for power supply lines
N/A
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential mode
±2kV common mode
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage DIPS,
short interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycle
<5% UT (>95% dip in UT)
for 5 sec
<5% UT (>95% dip in UT)
for 0,5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycle
<5% UT (>95% dip in UT)
for 5 sec
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the SonixTouch Q+
Ultrasound System requires continued operation during power
mains interruptions, it is recommended that the SonixTouch Q+
be powered from an uninterruptible power supply (UPS) or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
environment.
Note:
UT is the AC mains voltage prior to application of the test level.
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......