SonixTablet Service Manual
00.053.129, Revision C
Chapter 11: Maintenance
11-13
11.3.7 Sterilization
Sterilization of transducers is not possible. Follow the instructions for cleaning and disinfection instead:
• Endocavity transducers:
• Non-invasive transducers:
.and
.
11.4 SHIPPING TRANSDUCERS FOR SERVICE
It is the customer's responsibility to ensure:
• each transducer is disinfected prior to shipping (
• the transducer is properly packaged for shipment (
)
• all shipping waybills/paperwork is completed as per the relevant regulations and laws.
Refer to
7.6 Returning Parts for Service/Repair/Replacement
for more details on transducer return or replacement.
Note:
Where transducers (non-critical and semi-critical medical devices/equipment) cannot withstand
sterilization, the FDA recognizes the use of a sterile gel and a sterile transducer cover as an acceptable
method of infection control for ultrasound transducers.
Содержание SonixTablet
Страница 1: ...SonixTablet Ultrasound System Service Manual...
Страница 2: ......
Страница 4: ......
Страница 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Страница 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Страница 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Страница 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Страница 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Страница 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Страница 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Страница 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Страница 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...