5.4.8 "Input/Output settings" menu (if activated)
The input/output modules are:
As input:
o
Current,
o
Voltage,
o
Temperature (PT100/PT1000),
o
Contact.
As output:
o
Current/Voltage,
o
Relay.
The menu only appears if inputs or outputs are installed on the flow meter. To install additional
inputs/outputs, please contact Ultraflux to find out the specifications of all available inputs/outputs.
1) Current input and voltage input module
The possible options are:
OFF: deactivation,
ON: activation,
Simulation.
For ON and Simulation, the following must be defined:
the value corresponding to 4 mA (for a current input),
the value corresponding to 0 V (for a voltage input),
the sensor range,
the value to be simulated (in simulation mode),
the value of the filter and the memory (in ON mode).